PDS_VERSION_ID = PDS3
LABEL_REVISION_NOTE = "Initiated Unknown"
OBJECT = INSTRUMENT
INSTRUMENT_HOST_ID = P11
INSTRUMENT_ID = GTT
OBJECT = INSTRUMENT_INFORMATION
INSTRUMENT_NAME = "GEIGER TUBE TELESCOPE"
INSTRUMENT_TYPE = "ENERGETIC PARTICLE DETECTOR"
INSTRUMENT_DESC = "
Instrument Overview
===================
The University of Iowa experiment comprises seven miniature
Geiger-Muller tubes in various physical arrangements and with
various levels of shielding responding to electrons in several
energy ranges from 70 keV to tens of MeV, and protons in several
energy ranges from 700 keV to tens of MeV. No one of the individual
detectors distinguishes uniquely between protons or electrons. A
full analysis of the relative responses of the entire array of
detectors will provide separate energy spectra of electrons and
protons by using laboratory-determined unit response functions and
an identification matrix. Overall calibrations were provided by
observations during the outbound pass of Pioneer 10 through the
magnetosphere of the Earth on 3 March 1972 and during several solar
energetic particle events, especially those of July and August,
1972.
Principle Investigator : J.A. Van Allen
Build Date : 1973-04-06
Instrument Mass : 1.64 kg
Instrument Height : 0.145 m
Instrument Length : 0.171 m
Instrument Width : 0.145 m
Instrument Manufacturer : The University of Iowa
Instrument Serial Number : 0853-04
Scientific Objectives
=====================
Encounter Objectives
--------------------
The investigation objectives for the Pioneer 11 encounter data
(Jupiter, and Saturn) are:
1) The study of the intensity of planetary energetic particles as
a function of position in the magnetosphere. This includes a
determination of plasma boundaries that determine the structure
of the planetary magnetosphere.
2) The study of planetary energetic particle dynamics and
trajectories in order to better understand planetary magnetic
storms and plasma sources and sinks.
3) The study of low energy electron and proton events to
understand the magnetospheric propagation and acceleration of
such particles as a function of position.
Cruise Objectives
-----------------
The investigation objectives for the various Pioneer 11 cruise
data (Earth-Jupiter, Jupiter-Saturn, and Post-Saturn) are:
1) The study of the intensity of galactic cosmic rays as a
function of solar activity and as a function of heliocentric
position. Long-range objectives are to determine the position
of the modulation boundary of the heliosphere and the intensity
of the galactic cosmic rays in the interstellar medium beyond
it.
2) The study of solar energetic particles as a function of
heliocentric distance and their propagation in the
interplanetary medium.
3) The study of low energy electron and proton events to
understand the interplanetary propagation and acceleration of
such particles as a function of heliocentric distance.
Calibration
===========
The electrical power for the spacecraft is provided by four Type
SNAP 19 radioisotope thermal generators (RTG's) of the Atomic Energy
Commission. There are also several much weaker radioisotpe heater
units (RHU's) for spot heating. The heating of these systems is
provided by a mixture of plutonium isotopes; principally plutonium
238. Gamma and X-rays from the decay of some of the isotopes produce
a time varying background in the singles rates of the GM tubes.
These background rates can be calculated using the following
formulae.
R(A) = 1.137*RR
R(B) = 0.905*RR
R(C) = 0.958*RR
R(D) = 0.254*RR
RR = 1.103E - 02*[2*f(t)+0.9*f(t+0.832)]
+ 5.53E - 03*[3*g(t)+g(t+0.832)] counts/second
where
f(t) = 23.792*EXP(-0.3623*t) + 12.120*EXP(-0.0094*t)
- 35.912*EXP(-0.2432*t)
g(t) = EXP(-0.0081*t)
and t is measured in years with t=0 on March 15,1971.
[Calibration description exerpted from VANALLEN&RANDALL1985.]
Detector G is a thin solid state detector and is insensitive to this
radiation and hence requires no correction. Detector G also has a
small Am-241 alpha particle source at the edge of its viewing cone
to provide an inflight calibration. This source gives a background
counting rate of 0.06 counts per second.
Instrument Temperature
----------------------
The multiple GM detector rates and the solid state detector rate
require no correction as a function of temperature. The singles
rates of the GM tubes require a slight temperature correction.
This temperature correction is of the form, A' =A(T)*X(T) for
detectors A, B, and C, and for detector D, the form is
D'=D(T)*Y(T), where
X(T)=1.0+0.0003694*(75.0-T)
and
Y(T)=1.0-0.0006514*(75.0-T).
The temperatures in degrees F as a function of time for Pioneer 11
are given in the following table.
--------------------------------------------------------------
Pioneer 11 Temperatures
--------------------------------------------------------------
YEAR PERIOD (DOY) TEMPERATURE PERIOD TEMPERATURE
--------------------------------------------------------------
1973 96 77.8 97 67.0
98 64.8 99 65.8
100 67.0 101-119 69.9
120-125 67.0 126-131 64.8
132-148 58.1 149-153 55.9
154-205 50.1 206-226 46.6
227-272 50.1 273-312 48.4
313-317 43.7 318-365 46.6
1974 1- 28 46.6 29-290 44.7
291-365 42.9
1975 1-101 42.9 102-139 35.8
140-365 42.9
1976 1-167 42.9 168-366 41.1
1977 1- 24 41.1 25-302 39.29
303-365 37.52
1978 1-252 37.52 253-365 35.77
1979 1-149 35.77 150-365 34.03
1980 1-155 34.03 156-172 32.30
173-188 34.03 189-366 32.30
1981 1-216 32.30 217-365 30.58
1982 1-120 30.58 121-365 28.88
1983 1- 48 28.88 49-332 27.20
333-365 25.52
1984 1- 60 25.52 61 23.86
62 22.22 63-184 20.59
185-276 18.98 277-366 17.38
1985 1- 35 17.38 36-116 15.80
117-205 14.22 206-365 12.66
1986 1- 63 11.11 64- 70 9.55
71-139 8.00 140-310 6.43
311-365 4.86
1987 1- 31 4.86 32-189 3.22
190-304 1.63 305-365 -0.03
1988 1- 87 -0.03 88-231 -3.54
232-366 -5.33
1989 1- 7 -5.33 8-150 -7.12
151-292 -8.91 293-365 -10.70
1990 1- 70 -10.70 71-212 -12.49
213-355 -14.28 356-365 -16.07
1991 1-133 -16.07 134-275 -17.86
276-365 -19.65
Detectors
=========
Six miniature Geiger-Mueller (GM) tubes are used as basic detectors.
Three of these (A, B, and C) are EON Corporation end-window type
6213 tubes. The three other tubes (D, E, and F) are EON Corporation
type 5107 tubes. The six tubes were placed in a variety of physical
arrangements. The seventh detector (G) is a thin (28 micron thick),
single element solid state detector with a physical collimator. This
detector was made by Nuclear Diodes, Inc.
Tubes A, C, and B are mounted in a single block. The central tube C
is shielded omnidirectionally. Tubes A and B are similarly shielded
except for thinner window unidirectional collimators in the +X (s/c
coordinates) direction. The individual counting rates of the three
tubes are telemetered separately; also, double coincidences AB and
triple coincidences ABC with a resolving time of 1 microsecond are
formed and telemetered.
The second assembly comprises an omnidirectionally shielded
triangular array of three 5107 tubes. The rate of D and the triple
coincidence rate of the DEF are telemetered.
The third assembly uses the collimated solid state detector as input
to a charge sensitive amplifier. This signal is delay-line clipped
to 180 nanoseconds, amplified and then fed into a fixed level
threshold discriminator. This threshold corresponds 0.61 MeV for
protons. The output of the discriminator is then fed into an output
circuit which shapes the pulse so that it is similar to that from
the GM tubes; it is then treated in the same manner. Detector G
looks in the +X direction.
---------------------------------------------------------------
University of Iowa Geiger Tube Telescope Energy Ranges and
Geometric Factors of Pioneer 11 Detectors
---------------------------------------------------------------
Effective
Inverse
Omnidirectional 1
Geometric -------
Effective Energy Factor 4*pi*Q
Detector Range, MeV (1/Q),cm**-2 Type (cm*cm*sr)**-1
---------------------------------------------------------------
Electrons
A-C 0.04❮E❮21 730 D 58.1
B-C 0.56❮E❮21 785 D 62.5
G insensitive ... .. ...
to electrons
C E❯21 23 O ...
D E❯31 63 O ...
AB E❯21 6910 D 550
ABC E❯21 6910 D 550
DEF E❯31 3150 O ...
Protons
A-C 0.61❮E❮80 650 D 51.7
B-C 9❮E❮80 650 D 51.7
G 0.61❮E❮3.41 285 D 22.7
C E❯80 8.2 O ...
D E❯80 23 O ...
AB E❯130 2463 D 196
ABC E❯130 2463 D 196
DEF E❯150 11500 O ...
_________________________________________________________________
Type D=Directional O=Omnidirectional
Electronics
===========
The GTT electronics system consists of two basic sections. The first
is the power converter which regulates and filters the 28 volt, 20
kHz spacecraft power supply. The on-off command functions through a
solid state switch by removing power from the driving circuitry.
Output voltages of +7.75, +5.00, -12.00 and +900 are supplied to the
experiment. The 7.75 and 5.00 volt lines are regulated to 1% and the
900 volt lines are regulated with VR tubes.
The second section is the signal processor. This is used to
condition the data from the seven detectors and consists of a
MOS/TTL/transistor hybrid system containing MOS/TTL logic and
transistorized discrete component interface circuits. The processor
is completely redundant with the exception of the interface
circuits. Upon command to the spacecraft, the signal processor can
be switched from the main logic system to a stand-by redundant logic
system. The function of the processor is to sequentially accumulate
data on a frame basis from the seven detectors. Data are accumulated
in a 24 bit register and then compressed quasi-logarithmically to 12
bits for transmission.
Mounting
========
The Z-axis of the instrument is parallel to the axis of rotation of
the spacecraft. The +X axis points outward into free space from the
rim of the instrument compartment of the spacecraft. The
magnetometer and a portion of the magnetometer boom subtend a
trivial fraction of the fields of view of the collimators of A, B,
and G; otherwise, there is no physical obstruction within the fields
of view.
The rotational axis (+Z) of the spacecraft is pointed continuously
at the earth within an error of less than one degree and therefore
lies approximately in the ecliptic plane. The spin period and
sampling period are asynchronous; thus angular distributions of
particle intensities as a function of roll angle in the equatorial
plane of the spacecraft are assembled as a software operation by
using attitude data supplied by the Ames Research Center. In our
analysis the roll angle is measured from the ascending node of the
spacecraft's equator on the ecliptic to the +X axis of the
instrument at the midtime of the sample.
Operational Considerations
==========================
The instrument uses 12 bits in each 192-bit main science frame of
the spacecraft's telemetry format. Quasilogarithmic data compression
is used to maintain 1% accuracy at all possible counting rates. All
outputs are digital. A complete cycle of the GTT data comprises
eleven main science frames as follows: sync word, G, A, B, G, AB,
ABC, C, D, ABC, and DEF.
Counts from each detector channel are accumulated for a period of
time in seconds equal to 192/b, where b is the telemetry rate in
bits per second for the entire spacecraft (b=16, 32, 64, 128, 512,
1024, or 2048, as selected by ground command).
Operational Modes
=================
INSTRUMENT MODE ID : ON
GAIN MODE ID : N/A
DATA PATH TYPE : REALTIME
INSTRUMENT POWER CONSUMPTION : 1.6 ???????
The GTT experiment utilizes two commands. A power ON/OFF command and
a function command for Main/Standby processor selection.
[VANALLENETAL1980]
Section ID: 'GEIGER TUBES A, B, C, & D'
---------------------------------------
Scan Mode ID : N/A
Data Rate : 16-2048
Sample Bits : 24
Total FOVS : 1
Fields of view
FOV Shape Name : Cone
Horizontal Pixel FOV : N/A
Vertical Pixel FOV : N/A
Horizontal FOV : 30
Vertical FOV : 30
Parameters
Instrument Parameter Name : Particle Count Rate
Minimum Instrument Parameter : 0.1
Maximum Instrument Parameter : 2.0E04
Noise Level : 0.4
Instrument Parameter Unit : counts/second
Sampling Parameter Name : TIME
Minimum Sampling Parameter : N/A
Maximum Sampling Parameter : N/A
Sampling Parameter Interval : 115.5
Sampling Parameter Resolution : 0.75 ????????
Sampling Parameter Unit : second
Detector ID : Geiger Tube G
Electronics ID : GTT
Filter Number : N/A
Telescope ID : N/A
Section ID: 'GTT'
-----------------
Scan Mode ID : N/A
Data Rate : 16-2048
Sample Bits : 24
Total FOVS : 1
Fields of view
FOV Shape Name : Cone
Horizontal Pixel FOV : N/A
Vertical Pixel FOV : N/A
Horizontal FOV : 40
Vertical FOV : 40
FOVS : 1
Parameters
Instrument Parameter Name : Particle Count Rate
Minimum Instrument Parameter : 0.1
Maximum Instrument Parameter : 2.E04
Noise Level : .06
Instrument Parameter Unit : counts/sec
Sampling Parameter Name : TIME
Minimum Sampling Parameter : N/A
Maximum Sampling Parameter : N/A
Sampling Parameter Interval : 115.5
Sampling Parameter Resolution : .75 ???????
Sampling Parameter Unit : second
['Detectors', 'Electronics', 'Mounting', and 'Operational
Considerations' descriptions exerpted from VANALLENETAL1980.]"
END_OBJECT = INSTRUMENT_INFORMATION
OBJECT = INSTRUMENT_REFERENCE_INFO
REFERENCE_KEY_ID = "FIMMELETAL1977"
END_OBJECT = INSTRUMENT_REFERENCE_INFO
OBJECT = INSTRUMENT_REFERENCE_INFO
REFERENCE_KEY_ID = "VANALLENETAL1974"
END_OBJECT = INSTRUMENT_REFERENCE_INFO
OBJECT = INSTRUMENT_REFERENCE_INFO
REFERENCE_KEY_ID = "VANALLENETAL1980"
END_OBJECT = INSTRUMENT_REFERENCE_INFO
OBJECT = INSTRUMENT_REFERENCE_INFO
REFERENCE_KEY_ID = "VANALLEN&RANDALL1985"
END_OBJECT = INSTRUMENT_REFERENCE_INFO
END_OBJECT = INSTRUMENT