Trigger 12 Description
Sequence: All
Trigger 12 (Basic Survey Mode)
This mode is designed to provide a Survey with fast temporal resolution
and standard spectral resolution for periods when an IEB has not been loaded.
Periodic Waveform measurements are made. The entire RPWS frequency range
(1 Hz - 16 MHz) is covered. The onboard dust detection algorithm is activated,
and looks for dust using the LP sensor.
Receiver |
Frequency Range |
Sensors |
Time between Snapshots |
LFR |
1 - 25 Hz |
EX, BZ |
16 seconds |
MFR |
25 - 12 KHz |
EX, BX |
32 seconds |
WFR |
.05 - 2.5 KHz |
EX, EW, BX, BY, BZ |
341 seconds |
HFR |
3.6 KHz - 16 MHz |
EX, EW |
9.875 seconds |
The HFR is in CMS mode (Composite Mode Survey). No direction-finding is done.
This mode is defined by:
Correlations Channels Integ. Size Frequency
Band Ant. Auto- Cross- per Band df Period Rep Steps kHz Start - Stop
---- ---- ----- ------ -------- -- ------ --- ----- ---- ------------
ABC 2E y y 16 n 1000 1 3.6-319 kHz
H1 2E n y 2 n 80 1 60 25 325k-1.8MHz
H2 1E n n 1 n 20 1 72 200 1.8M-16MHz
Bands ABC require 3.16 sec to complete, H1 requires 5.05 sec, and H2 requires 1.46 sec
for a total of 9.67 sec per sweep. Data collection and output limits the snapshot
time to 9.875 seconds.